Tympanometer – Tymp 4000-M

The HOMOTH Tympanometry PC module Tymp 4000-M conducts fully automatic impedance measurement within seconds, with all functions, curves, and data displayed on a PC monitor. This ensures viewing of results before documentation, avoiding incorrect printouts. Ideal for routine examinations and clinical diagnosis, it features a quick compliance measurement, completed in two seconds, suitable for children and restless patients. Its objective results are independent of patient involvement. With simple one-handed operation, the probe is positioned near the external auditory canal, with a two-color LED indicating correct placement and automatic start of the measurement process. All data and curves are electronically saved for integration into practice management software.

Technical Specifications

  • System: PC module
  • PC requirement: Pentium PC from 500 MHz, USB 2.0 port, Windows XP, 7, 8, 10 (32/64 bits)
  • Measurement method: Impedance and reflex measurement
  • Measuring tone: 226Hz, 85dB SPL
  • Print area: +200 to -400 daPa
  • Pressure change: 300 daPa/sec
  • Measuring time: 2 sec. for compliance
  • Reflex tones: 500, 1000, 2000 and 4000 Hz at 85, 95 and 105 dB SPL
  • Reflex detection: automatically
  • Reflex measurement: ipsi-lateral and optionally contralateral
  • Pressure calibration: automatically when dialed
  • Status display: three-color LED in the probe, detailed display symbols
  • Dimensions: 330 x 80 x 265 mm (WxHxD)
  • Power supply: 13.2 V, ext. Power supply 100-240V, 50/60Hz, 30W
  • Accesories: Set of earplugs, probe holder, headphones DD 45 C (with contraOption), USB cable, power supply, installation CD, operating instructions